Datex-Ohmeda Biox 3740 Pulse Oximeter
Item #: 0473MDC
The Datex-Ohmeda Biox 3740 oximeter is a standalone, noninvasive, arterial oxygen saturation monitor. It provides continuous, real-time Sp02 level, a pleth waveform and pulse rate readings. Trend information is available through both the analog and digital output ports. It is similar to the model 3700 only in a smaller package. It can operate off a rechargeable internal battery and comes complete with a new re-usable finger sensor.
Product Description:
- Reconditioned units
- 90-day warranty
- New Finger Probe (Reusable Adult)
- 110 volt cmpatible 60 Hz
- Excellent condition
Features:
- Alarm Silence Key: silences all audible alarms for 120 seconds - Two Displays: Switches between Large Waveform and Large Number display for easy reading
Menu/Enter Key: Makes it easier to toggle between menus
- Up and Down Keys: Conveniently adjusts pulse volume during monitoring without using menu
- Alarms: Low Quality Signal Indicator, Low Battery Indicator, High Sp02 and Low Sp02 Limits
- Functions with Ohmeda Probes (Finger Sensors)
Specifications:
Dimensions:
- Size (W x H x D): 8.0in x 2.8in x 8.8in
- Weight: 5.5 lbs.
- Clock: 24-hour clock
Power Ratings:
- Power Consumption: 30 volt-ampere maximum
- Wall Mount and Table Transformer
Primary: 120 ±10% VAC 50/60 Hz 100 ±10% VAC 50/60 Hz 220/240 ±10% VAC 50/60 Hz Secondary: 15.5 VAC at 1.4 ampere
- Battery: Internal 12V, 1.9 ampere-hours sealed lead-acid. Continuous charge while operational
- Charge Time: 100% capacity - 6.5 hours
- Overload Protection: 2 ampere fuses
Operational Ratings:
- Saturation Range: 0 to 100%
- Saturation Accuracy1 standard deviation
90 to 100%: 1.5% 80 to 89.9%: 2.1%
- Overall Range 60 to 100%: 2.4%
Below 60%: Unspecified
- Pulse Rate Range: 40 to 235 bpm
- Pulse Rate Accuracy: ±1.7% of current reading at constant pulse rate
Operational:
- Display: LCD, selectable numeric or waveform emphasis
- Signal Validation: Signal strength indicator and waveform plethysmographic display for signal validation
- Memory: 8 hour data storage for SpO2 and pulse rate
- Saturation Calculation: 30 values per second at 60 Hz, 25 values per second at 50 Hz
- SpO2 Response Mode: Selectable 3, 6, 12 seconds (preset 6 seconds)
- Diagnostics: Automatic self-test on start-up, extensive service diagnostics
- Outputs: Digital (RS-232C) and analog outputs
Alarm:
- Alarm Limit Range
Low SpO2: 50% to 100%(preset 90%) High SpO2: 7% to 100% (preset OFF) Low Pulse: 40 to 200 bpm (preset OFF) High Pulse: 70 to 250 bpm (preset OFF)
- Alarm Silence: 2 minute silence (automatical on start-up)
|